Data Recording has
never been
this easy

A whole new level of simplicity makes set-up and operation faster and more intuitive than ever before. MULTI-TOUCH gestures like pinch and zoom make the TrendCorder as easy to use as your smart phone. Finally, an amazing chart recorder replacement with world-class DEWETRON precision and reliability - at a great price!

  • REALLY EASY TO USE
    Start recording by one touch, and then enjoy the intuitive 64 bit application with multi-touch technology to browse and zoom your data. Gestures that you use on your smart phone, also work on the TrendCorder. Figure out the power of DEWETRON´s special DejaView™ feature.
  • FLEXIBLE & EXPANDABLE
    There are 4 slots for user exchangeable high-speed TRION™ modules for virtually every sensor. Low-speed inputs, like thermocouples, connect directly to the EPAD connector, and do not use any of your high-speed slots. Add a DEWE2-F chassis to get more high-speed slots at any time.
  • MOBILE AND RUGGED
    This instrument is designed in a nice briefcase format, weighs only 18.7 pounds with 18 channels installed and is supplied with a handy carrying bag. It is shock and vibration tested against really tough standards. Get your own TrendCorder and challenge it´s ruggedness.
  • WHAT A GREAT VALUE
    If it weren't for the 1500V of galvanic isolation that each of the TrendCorder's 18 channels provides, you might be "shocked" by its low price. This includes 18 250 kS/s/ch input channels, a crisp multi-touch display, the easiest to use acquisition software around, 1 TB of removable data storage, and a carrying bag.
  • Measure

    After switching the TrendCorder on it´s immediately acquiring data, recording can be started by a single touch. Basic measurements don´t take more than start – stop – view. Of course, freely configurable screens, triggers, markers, cursors, etc. are available, too. The integrated multi-touch support simplifies browsing and zooming data enormously.

    Comparing - Analysing - Reporting

  • DejaView™

    To keep an overview of your current test at all times, we have invented the special DejaView™ feature. It allows you to view all data in the past - since start of recording - and is especially useful for long term testing. Live data can still be seen at the same time on a different recorder, on the same or on a different screen.

  • Reporting

    Comparing data sets and performing basic analyses like statistics are easily possible. A layout for a report can be created by simply dragging signal curves, lettering, shapes and a logo onto a print-space.

  • Measure
  • DejaView™
  • Reporting

Data Acquisition Modules

TRION™ modules combine the power of simultaneous sampling ADC technology with advanced DEWETRON Signal Conditioning - in one convenient module that plugs right into your TrendCorder. Modules are available with 16 or 24 bit resolution and sampling speeds of up to 2 MS/s per channel, and most modules provide high isolation. High accuracy is guaranteed and the self-test function ensures confidence before important measurements. These modules give you the flexibility you need for changing tasks and the performance for challenging measurements.

TrendCorder-18 TrendCorder-6
Input channels 18 6
Input specification Isolated voltage and current (via ext. shunt) inputs with 1500 V isolation 16-bit resolution, 250 kS/s per channel, ADC per channel Freely programmable 2nd, 4th, 6th and 8th order filters and ranges from ±5 mV to ±100 V Self-check functionality and detailed DEWETRON factory calibration report included
Open slots for TRION™ modules 1 3
High speed expansion Prepared to connect one or more front-end chassis at any time by PCIe and SYNC interfaces
Low speed expansion 12 Hz EPAD2 interface included for external Low speed modules
TrendCorder-18 TrendCorder-6
PC configuration 2.5 GHz Intel® Core™ i5 processor, 8 GB RAM, bright 15.4” multi-touch display 6 x USB, 2 x LAN Ethernet, 1 x external monitor, 64-bit Microsoft® Windows® 7 operating system (Linux OS optional)
Data storage 1 TB HDD dedicated for data storage (upgrade to 480 GB SSD available) 120 GB SSD for operating system and application software, both in a single removable drive bay
Power supply (max.) 90 to 264 VAC
Dimensions (W x D x H) 462 x 320 x 135 mm (18.2 x 12.6 x 5.3 in.)
Weight incl. modules 8.5 kg (18.7 lb.) 7.9 kg (17.4 lb.)
Power consumption incl. modules Typ. 120 W Typ. 100 W
TrendCorder-18 TrendCorder-6
Operating temperature 0 to +50 °C, down to -20 °C with prewarmed unit
Storage temperature -20 to +70 °C
Humidity 10 to 80 % non cond., 5 to 95 % rel. humidity
Max. altitude 3000 m (10000 ft)
Sine vibration1) (EN 60068-2-6) Acceleration 20 m/s², Freq. 10 Hz - 150 Hz, Sweep 1 oct/min, 20 cycles
Shock1) (EN 60028-2-27) Acceleration 30 g, duration 11 ms, pulse form half sine, 3 pumps/direction, 6 directions
Random vibration1) (EN 60721-3-2) Class 2M3 (spectral acceleration density 3 m²/s³, frequency range 10 Hz-200 Hz, duration 30 min/direction)
TRION-1603-LV - isolated voltage input module
Input channels 6 using BNC sockets (TRION-1603-LV-6-BNC)
Sampling rate 250 kS/s per channel
Resolution 16 bit
Input ranges ±5, ±10, ±20, ±50, ±100, ±200, ±500 mV, ±1 V, ±2 V, ±5 V, ±10 V, ±20 V, ±50 V, ±100 V
Input configuration Isolated
Input coupling DC
Overvoltage protection ±300 VDC
Bandwidth (-3 dB) 100 kHz
Power consumption 6 W
TRION-1620-LV - isolated voltage input module
Input channels 6 channels BNC; voltage input;
Input ranges Voltage ±5 mV to ± 100 V)
Sampling rate >1 to 2 MS/s, 16-bit
100 S/s to 1MS/s, 24-bit
Input configuration Isolated
Isolation 1.5 kV
Overvoltage protection ±300 VDC
Bandwidth (-3 dB) 1 MHz 2nd order Bessel filter
Power consumption 6 W
TRION-1620-ACC - isolated voltage and IEPE input module
Input channels 6 channels BNC; voltage input; IEPE®; 1 counter
Input ranges Voltage ±5 mV to ± 100 V1)
IEPE ±5 mV to ±5 V, excitation 4 or 8 mA
Sampling rate >1 to 2 MS/s, 16-bit
100 S/s to 1MS/s, 24-bit
Input configuration Isolated
Isolation 1.5 kV
Overvoltage protection ±300 VDC
Bandwidth (-3 dB) 1 MHz 2nd order Bessel filter
Counter 1 x counter channel linked to analog channel #1; trigger level 70 % of actual analog input range
Counter modes Event counting, period, frequency, pulse width, duty cycle
Power consumption 6 W
TRION-2402-V - isolated voltage (and current) module
Input channels 8 (TRION-2402-V8-B), double slot width connector panel
4 (TRION-2402-V4-B ), single slot width connector panel
Sampling rate 204.8 kS/s/ch simultaneous sampling
Resolution 24-bit, with anti-aliasing filtering
Input ranges Voltage: ±300 mV to ±1000 V
Current: depending on shunt connected (external); e.g. 20 mA, 1A, 5A
Bandwidth (-3 dB) 77 kHz
Measurement category CAT II 300 V
Isolation voltage 4 kV
Non-destructive voltage Range ≤ 10 V: ±1000 VDC
Range > 10 V: ±2000 VDC
Connector 4 mm safety banana sockets
TRION-2402-dSTG - differential universal input module
Input channels 8 using RJ-45 sockets (TRION-2402-dSTG-8-RJ)
6 using LEMO 1B sockets (TRION-2402-dSTG-6-LEMO)
Sampling rate 204.8 kS/s per channel
Resolution 24-bit
Input ranges Voltage: ±10 mV to 10 V
Bridge: 1 to 1000 mV/V or mV/mA
IEPE: ±100 mV to 10V
Resistance: 10 Ω to 30 kΩ
Current: depending on external shunt
Input configuration Single ended or differential (programmable)
Input coupling DC, AC (0.16 Hz, 0.5 Hz, 3.4 Hz, 10 Hz)
IEPE Excitation 4 mA ±10 %
Supported sensors 4- or 6-wire full bridge
3- or 5-wire ½ bridge with internal completion (software programmable)
3- or 4-wire ¼ bridge with internal resistor for 120 and 350 Ω (software programmable)
4-wire full bridge with constant current excitation (piezoresistive bridge sensors)
Potentiometric; resistance
Pt100, Pt200, Pt500, Pt1000, Pt2000 (software linearization)
Shunt calibration Two internal shunt resistors 50 kΩ and 100 kΩ
Automatic bridge balance ±250 % of range
Bandwidth (-3 dB) 77 kHz
Teds support Yes
TRION-2402-dACC - differential multifunction input module
Input channels 8 with SMB sockets (TRION-2402-dACC-8-SMB)
6 with BNC sockets (TRION-2402-dACC-6-BNC)
Sampling rate 204.8 kS/s/ch simultaneous sampling
Resolution 24 bit
Input ranges Voltage: ±30mV to ±100 V
IEPE: ±100 mV to 10 V
Resistance: 10 Ω to 1 MΩ
Current: depending on external shunt
Input configuration Single ended or differential (programmable)
Input coupling DC, AC (0.16 Hz, 0.5 Hz, 3.4 Hz, 10 Hz)
Excitation current 0.1 to 24 mADC (programmable, 16 Bit DAC, 2 ranges)
Supported sensors IEPE (up to 24 mA excitation)
Resistance
Resistance Temperature Detection: Pt100, Pt200, Pt500, Pt1000, Pt2000
Counter modes Event counting; periode; frequency; pulsewidth; dutycycle
Trigger level Trigger and retrigger level free programmable within analog input range
Bandwidth (-3 dB) 77 kHz
Over voltage protection ±100 VDC (200 Vpeak) input to GND
Teds support Yes
TRION-BASE - Counter, DIO and synchronization module
Counter channels 2 SuperCounters™
Basic event counting, gated counting, up/down counting and encoder mode (X1, X2 and X4)
Period, frequency, pulse width, duty cycle and edge separation
Encoder (angle and linear), gear tooth with/without zero, gear tooth with missing/double teeth
Digital I/O 8 DIO and 8 DI
CMOS/TTL
TTL, 20 mA
±30 V permanent, 50 peak (for 100 msec)
Synchronization IRIG code B, DC I/O and Sync I/O
Isolated 350 VDC
Additional features 1 AUX socket (by default set to camera trigger)
Camera trigger, trigger input/output, acquisition clock and programmable clock output
TRION-CNT - isolated COUNTER/ENCODER interface module
Input channels 6 counters or 18 digital inputs (TRION-CNT-6-LEMO)
Counter modes:
Event counting Basic event counting, gated counting, up/down counting and encoder mode (X1, X2 and X4)
Waveform timing Period, frequency, pulse width duty cycle and edge separation
Sensor modes Encoder (angle and linear), gear tooth with/without zero, gear tooth with missing/double teeth
Rated input voltage (max.) 70 VDC (46.7 Vpeak)
Compatibility Adjustable trigger levels
Configuration isolation (±500 VDC) for each input channel
Input coupling DC and AC (1Hz) AC for input A only
Bandwidth (-3dB) 5 MHz
Trigger adjustment range 0 to 50 V
Overvoltage protection ±100 V continuous
Counter resolution 32-bit
Counter time base 80 MHz
Maximum input frequency 10 MHz
Sensor power supply (per module) 5 V (600 mA) and 12 V (600 mA), not isolated
TRION-DIO-4800 - isolated digital input module
Input channels 48 isolated digital inputs (TRION-DIO-4800)
Compatibility CMOS
Configuration Isolated input
Input low level UIN < 1.8 V
Input high level UIN > 3.2 V
Bandwidth 3 MHz
Overvoltage protection 35 V continuous (65 V peak)
Isolation voltage (channel to channel) 100 V
Isolation voltage (input to output) 250 V
DEWE2-F4s

Instrument facts

  • 4 slots for TRION™ modules
  • 90 to 264 VAC
DEWE2-F7s

Instrument facts

  • 7 slots for TRION™ modules
  • 90 to 264 VAC
DEWE2-F13s

Instrument facts

  • 13 slots for TRION™ modules
  • 90 to 264 VAC
DEWE2-F18

Instrument facts

  • 18 slots for TRION™ modules
  • 90 to 264 VAC
EPAD2-TH8-x

Instrument facts

  • 8 isolated thermocouple inputs
  • Thermocouple type K, J, T, E, R, S, B, N, C, U
  • 350 V isolation

Information